NeuroTrac MyoPlus2 with Software
This unit is a dual channel EMG (Biofeedback) combined with NMS-STIM (Neuromuscular Stimulation) & ETS (EMG Triggered Stimulation)
Item No.: ERP1387-00
Request a Quote
Special pricing for Healthcare Professionals | Login to view
Professionnels de la santé | connectez-vous pour prix spéciaux
Latex Warning
Additional Shipping Charges
- Passive and Active phases: EMG and NMS in one programme.
- Manual and Automatic threshold control.
- EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
- Stand alone device or used in conjunction with established multilingual software to be dowloaded communicating via Blutooth
- Pre-set neuromuscular Stimulation modes for: Continence, Sports, Rehab.
- Custom programmes enabling up to 5 phases (any of: STIM W/R-MODCONT- ALT, EMG, ETS) to be set up in each programme.
- Multilingual LCD screen and voice prompts for Biofeedback.
- For home users: Produces and prints (with the PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation. - 4 x AA batteries
- Now complete with Software
Features of the software: - Connected to computer via dongle Blutooth
- An easy to use windows software used to display live EMG readings on to your computer screen. Being able to view the EMG reading in a graph format makes it much easier for the therapist and patient to understand muscle activity.
- The software offers five main modes of operation including open display used for general EMG readings without a set time limit. Template training allows a therapist to create a template line for the patient to follow by increasing the patient’s control of a particular muscle. Work/rest assessment monitors the progress of the patient and creates statistics to aid the therapist to evaluate and diagnose muscular problems. Full EMG statistics are also displayed at the end of the session. Patient progress reports can be created from work/rest assessments and produce comprehensive graph and statistical reports.
- Up to 10 assessments can be viewed in graph form on one page allowing the therapist to visually compare each assessment to see what progress the patient is or is not achieving. These reports can also be used to form comprehensive reports if a clinic needs to demonstrate to a third party that the patient is benefiting from treatment.
The software requirement as follow:- PC computer with Internet connectivity for dowloading software and licence activation
- OS recommended Windows Vista / 7/8 / 10. Windows 10.
- Monitor 800 X 600 (minimum).
- 5 Go hard drive space.
- Apple Mac, iPad, Windows RT are not supported
TECHNICAL INFORMATIONS:EMG (Biofeedback) - Dual channel EMG
- EMG Range: 0.2 to 2000 μV RMS (continuous)
- Sensitivity: 0.l μV RMS
- Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
- Selectable Bandpass filter – 3db Bandwidth,
- a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% – Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% – Reading above 235 microvolts
- b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- Notch filter: 50 Hz (Canada 60Hz) – 33 dbs (0. 1% accuracy)
- Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- Work / Rest periods: 2-99 seconds
- Number of Trials: 1-99
STIM (Neuromuscular Stimulation) - Dual channel Stimulator
- Amplitude: 0-90 mA into 500 Ohm load – actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.
- Type: Constant current, maximum output voltage 70 Volts +5 / -10 Volts
- Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current
- Pulse width selection: 50-450 μS (2% accuracy)
- Pulse rate selection: 2-100 Hz (2% accuracy)
- Work / Rest periods: 2-99 seconds
- Time 1 – 99 minutes
- Ramp up time: 0.1 – 9.9 seconds
- Preset and user programmable treatment Programs
- Automatic output shut off with detection of open electrode above 0.5 m